Dimensions::
|
see dimension drawings
|
|
Mains voltage:
|
230 V/50 to 60 Hz tolerance +/-10% 115 V/50 to 60 Hz tolerance
+/- 10%
|
|
Power consumption:
|
400 VA during heating up 100-200 VA during operation
|
|
Series fuse recommended:
|
10 A
|
|
Ambient temperature:
|
-20 °C to +55 °C other temperatures on request
|
|
Immunity from disturbance:
|
according to EMVG and
Low Voltage Directive 72/73 EWG EN 50081-2: July
1993 EN 50082-2: March 1995
|
|
Relay outputs: floating:
|
230 V~ 5 A resistive load
|
|
Analogue input of the cell::
|
Re: >9 MOhm Ue: -4,5mV to
+265 mV
|
|
Resolution of A/D converter in the active measuring range:
|
14 Bit + sign
|
|
Analogue input of thermocouple:
|
Re: >900 kOhm
|
|
Temperature compensation:
|
electronic
|
|
Signal output 0/4 to 20 mA:
|
burde max. 500 ohm, potential-free
|
|
Response time of mA output:
|
< 200 ms
|
|
Display:
|
LCD, LED-illuminates, 240 x 64 points, graphic display
|
|
Interfaces:
|
RS 232, RS 485, bus compatible
|
|
Accuracy:
|
+/-0,2% from the measured value
|